GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 79
Reference Calibration is used to re-adjust either the integrated
front sensor of a display or a MD-N2M5B mounted on top of the
display. To re-adjust a display sensor, both the display sensor and
an external sensor need to be connected. Click the Reference
Calibration button and perform the re-adjustment by following
the displayed messages.
NOTES:
The Reference Calibration button is inactive if the selected display either does not
have a front sensor or the serial number of a MD-N2M5B sensor was not selected.
MD212G3/MD215MG/MD211G5/MD302C6 models do not support Reference
Calibration! The Reference Calibration button is inactive (grayed out) if one of these
models is selected. Therefore, please ignore the message “Reference Calibration is
highly recommended” for these models.
Instead, please select External Sensor as the Preferred Sensor. The integrated front
sensor of these models is re-adjusted automatically during a calibration using an
External Sensor.
Luminance Offset: An integrated front sensor or MD-N2M5B sensor can only
measure luminance in the periphery of the display screen surface. However, due to
the characteristics of an LCD display, there are differences in luminance and color
between the central region and the periphery of the display screen surface. When a
front sensor or MD-N2M5B sensor is being used, it is necessary to re-adjust such
differences (offset) in comparison with an external sensor. In case of integrated front
sensors, this offset is stored in the display itself. In case of the MD-N2M5B sensor, the
offset is stored in GammaCompMD QA Client.
Special care using MD-N25B: After the OK button is clicked in the Reinitialize Display
Configuration dialog box, as shown in Figure 89, specify the serial number again. If
this causes the combination of the display and the MD-N2M5B sensor to change then
it is necessary to repeat the reference calibration.
Reference calibration cannot be done with an external sensor without color support.
Please also refer to Additional considerations - Using sensor models without color
measurement capability below for details.