GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 15
2.6. Using the IBA LXcan or LXchroma instrument
For ambient light measurement, the optional LxLs lux sensor needs to be attached to
the LXcan or LXchroma instrument. Please turn the LXcan or LXchroma power to OFF
when attaching / detaching the Lux sensor. Please refer to detailed instructions how to
use this instrument in the LXcan or LXchroma user manual.
The screen contact mask is needed for measuring directly on screens.
The distance mode requires a distance of about 50 cm for measuring, an ultrasound
range finder is integrated in the LXcan or LXchroma. On the display, the distance is
shown as an arrow indicating in which direction the device must be moved to reach the
right measurement distance.
2.7. Using the Unfors Luxi (new product name: Raysafe Solo Light) instrument
It is equipped with a light detector which can be used for both measuring display
luminance and ambient light. Please turn the sensor's power OFF when enabling /
disabling the ambient light detector. Be careful to not pull out the USB cable when
performing USB communications with a USB - serial conversion cable.
NOTE: Only Unfors Luxi instruments equipped with firmware version 5.05 or later are
supported by GammaCompMD QA Client software. Please refer to detailed instructions
in the Unfors Luxi (XI Kit) User Manual.
2.8. Using the Windows Power Management Option
When the power management option is used in Windows XP (or when the Microsoft
Windows PC Automatic Energy-Saving Programis used), an external sensor may not
be recognized after the system returns from standby or sleep mode. If a sensor is not
recognized, remove it, reconnect it, and check that the external sensor automatic
detection and calibration are working normally.
2.9. Using Displays without PIP Support
With a display model not supporting PIP MODE, GammaCompMD QA Client cannot be
used when the display is connected using multiple inputs (e.g. DVI input + DisplayPort
input) through cables to one or more workstations. Please make sure that only one
input is connected to the display before using GammaCompMD QA Client. Also, please
perform a re-initialization of the display configuration when you disconnect/reconnect
display cables to change connection.