GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 104
NOTES:
Do not perform a Stand Alone Calibration when you are using the Main Screen of
GammaCompMD QA Client at the same. The results may be corrupted.
If the Stand Alone Calibration has a timestamp which is newer than the system time,
a warning dialog (Figure 87) is displayed when the Main screen of GammaCompMD
QA Client is opening. The results of the Stand Alone Calibration cannot be imported.
Repeat the Stand Alone Calibration and enter the correct time.
When Gamma Adjust is performed, both the result of the Gamma Adjust and the
result of the DICOM Measurement are imported into the database of
GammaCompMD QA. When DICOM Measurement is performed, only the result of
the DICOM Measurement is imported.
Figure 87: Warning Dialog