This Quick Reference Guide refers to the following NEC MDview Series 2
models
- MDview 193 19" Display, 1.3 mega-pixel = 1280 x 1024 resolution (L195RR)
- MDview 202 20" Display, 2.0 mega-pixel = 1600 x 1200 resolution (L205GR)
- MDview 213 21" Display, 2.0 mega-pixel = 1600 x 1200 resolution (L215GW)
- MDview 242 24" Display, 2.3 mega-pixel = 1920 x 1200 resolution (L246T0)
- MDview 242-2 24" Display, 2.3 mega-pixel = 1920 x 1200 resolution (L246T0)
- MDview 262 26" Display, 2.3 mega-pixel = 1920 x 1200 resolution (L266RZ)
- MDview 262-2 26" Display, 2.3 mega-pixel = 1920 x 1200 resolution (L266RZ)
Intended Use in Medical Environment
The MDview Series 2 colour displays are intended to be used for secondary diagnostics and review of medical images by
trained physicians.
Unpacking, Installation and calibration of this display should only be done by authorized and trained personnel. Any
installation done by a non-authorized person is done under his own risk and we accept no responsibility in any device
malfunctioning.
WARNING: - MDview Series 2 Displays are not certified as Medical Devices according to EU Directive 92/43/EC.
- DO NOT use 193, 202, 213, 242 or 262 display in patient vicinity.
- Any use for primary diagnostics is the sole responsibility of the hospital or clinic operator only.
MDview Series 2 Displays may be used as a component within a medical system or a PACS diagnostics workstation. Any
required re-certification process to comply with the EU directive 92/42/EC for medical devices is however within the
responsibility of the system/workstation supplier.
MDview Series 2 enhanced functionality and performance
MDview Series 2 models come with a factory preset DICOM like gamma correction enabled. This option provides an
essential improvement for viewing DICOM compliant medical images. However, this option cannot completely substitute a
true DICOM Part 14 GSDF curve.
Internal programmable 12-bit LUTs allows true DICOM GSDF calibration via software and calibration instrument,
complying with the DICOM Part 14 standard.
Improved Spatial Uniformity using Digital Uniformity Control (DUC).
Meets the visual requirements of acceptance tests and constancy tests under DIN V6868-57 Category B.
Meets the visual requirements as “AAPM TG18 Secondary Display”.
Shipping content includes a Software/Utility CD with:
GammaCompMD Client calibration software.
NEC tested and approved display controller drivers.
MDview Series 2 Online User’s Guide - with extended setup information as PDF documents.
NOTE: To prevent any un-intentional de-adjustment of optimum settings and calibrated values, NEC Display Solutions
Europe highly recommends the activation of the “OSD LOCK OUT with no control” function within the display’s
On-Screen-Display (OSD). This operation is described on page 26 of this document.